Post Event Follow-Up/Troubleshooting

Use the tools on this page to assist you in post event follow-up for failures and not graded results. There are a variety of tools offered; so, you can choose what best fits your needs. 

Troubleshooting guides and worksheets

To view the Post Event Troubleshooting guide in full, click on the red button above.

What's included in the guide?

Take a look at what's included in the Post Event Troubleshooting guide below:

The tools within the Post Event Troubleshooting Guide are not intended to replace any internal protocols used by your laboratory but may be used to provide additional support in post event follow-up.

Interactive Follow-Up and Troubleshooting Table

Self-Assessment

Exception code entered instead of a result (100%)

Suspected Cause

Laboratory was not able to perform testing or submit results by the due date. An exception code and comment were submitted in lieu of results.

Troubleshooting Action

  • Once able, perform testing on your PT samples and self-evaluate your result performance vs. expected results found in the Statistics Report. Save with your PT Review peer data submitted with your Event Statistics to compare your results with other labs.
  • Any Comment submitted with Exception Code can be viewed on your Data Submission Report. Save DSR with your self-evaluation.

Corrective Action

  •  Documentation of your exception code, reason for use (in lieu of results) and any self-evaluation will be required for audits or inspections.
  • Save all documentation of your PT records.

Suspected Cause

PT Provider was unable to score submitted results.

  • Small peer size (<10 labs submitting results)

  • Lab’s instrument/method data is significantly different than other instrument/method data and could not be combined with any other composite scoring group.

Troubleshooting Action

  • Evaluate your lab’s reported results in comparison with the expected results/range on the Evaluation Report.

  • Review peer data submitted with your Event Statistics to compare your results with other labs.

Corrective Action

  • Document your performance and save with your PT records.

  • Consider changes to your testing method/instrument if aged, outdated, or obsolete.

  • Perform split sample comparisons with another lab that uses the same test method using patient samples. If possible, choose samples that are similar in concentration to the PT samples under review. This is another tool to assess accuracy of patient results.

Suspected Cause

Results submitted by all labs in the scoring group showed large variability and could not agree on a result to meet PT grading requirements.

Troubleshooting Action

  • Evaluate your lab’s reported results in comparison with the expected results/range on the Evaluation Report.
  • Review peer data submitted with your Event Statistics to compare your results with other labs.
  • Retest the PT samples if necessary. Retested results may still not match expected results.

Corrective Action

  • Review the Event Notes for a summary of any potential PT scoring issues.
  • Save your self-evaluation with your PT records.

Fail

No results received (0%)

Suspected Cause

Results were not submitted before the due date.

Troubleshooting Action

  • Perform testing on your PT samples and self-evaluate your result performance vs. expected results found in the Statistics Report. Follow up on any failures and document. Save with your PT records.

Corrective Action

  • Review your Data Submission Report (DSR) for any blank/missing results before the due date has passed – missing results display with dashes.
  • Add PT ship dates and due dates to laboratory calendar.
  • Set reminders to test PT samples and enter results before the due date.

Suspected Cause

Clerical Error

  • Transcription/Transposition
  • Decimal Error
  • Incorrect units
  • Calculation error
  • Incomplete data
  • Incorrect instrument/method/reagent
  • Reported under wrong analyte
  • Incorrect use of <,> symbols

Troubleshooting Action

  • Check that original printouts match the results on your evaluation report.
  • Check for PT results that could have been switched when entering results.
  • Check units of measure and calculations are correct.
  • Check that all samples were reported with a result or an exception code on your Data Submission Report.
  • Check that instrument/method/reagent is correct on your Evaluation Report and that it is evaluated against correct peer group.
  • Greater than > and less than < symbols should only be used when the result is above or below the instrument’s reportable range.

Corrective Action

  • Review PT reporting process and perform a careful review of PT results prior to submission including review of the Data Submission Report.
  • Review PT Central setup is correct BEFORE entering results. Contact PT with changes to instruments/methods/analytes/units/decimals. Certain instrument changes may require a program change.

Suspected Cause

Specimen Mix-up

  • Wrong samples used

Troubleshooting Action

  • Check the sample ID carefully and rerun the sample.

Corrective Action

  • Review sample identification processes and that the correct test is being performed.

Suspected Cause

Specimen Handling

  • Samples stored improperly
  • Delivery problem
  • Dilution/pipetting error
  • Time delay/testing sequence not followed
  • Improperly mixed
  • Misinterpretation of instructions

Troubleshooting Action

  • Check that samples were received at correct temperature, on time, and stored properly.
  • Check that pipetting and measurement devices are dispensing correctly – when were they last calibrated/are they damaged.
  • Avoid delays/interruptions during sample testing.
  • Verify that test procedures are up to date and are being followed.
  • Review PT sample instructions to ensure they were followed. PT samples may require different procedure than patient samples.

Corrective Action

  • Train staff on proper routing, storage (according to coversheet instructions) and handling of specimens.
  • Have master lab calendar of expected arrival of PT samples. Contact PT provider for tracking information if samples do not arrive on time.
  • Develop a policy for re-training, continuing education, and competency assessment.
  •  

Suspected Cause

Reagents

    • Reagent lot change
    • Kit/reagent recall
    • Improper storage of standard/reagent/kit
    • Near expiration or out-of-date standard/reagent/kit being used

Troubleshooting Action

  • Check if the reagent lot changed before PT samples were tested.
  • Review the last lot to lot comparison to determine if the new lot caused significant bias in results.
  • Check the QC, QA and Preventative Maintenance (PM) Records the day that PT was run to be sure daily maintenance was performed, kits/reagents were not expired and the QC was in range.
  • Contact the reagent/kit vendor to see if other labs have reported any problems.

Corrective Action

  • Ensure process for testing new reagent lots uses patient samples (instead of QC or calibrators) and has appropriate acceptance criteria that will detect lot to lot changes beyond defined acceptable limits.
  • Review lab processes for handling of reagents – both storage and expiration date management.

Suspected Cause

    • Instrument technical problem

Troubleshooting Action

  • Check to see if there is an action log indicating problems prior to or immediately after running PT samples.
  • Check the QC, QA and Preventative Maintenance (PM) Records the day that PT was run to be sure daily maintenance was performed, kits/reagents were not expired and the QC was in range.

Corrective Action

  • Determine if maintenance fixed problems identified by reviewing QC and PT results.
  • Contact manufacturer for troubleshooting assistance.

Suspected Cause

    • Calibration Issue

Troubleshooting Action

  • Check SDI values for the analyte. Are SDI values all negative or all positive, +/-2?
  • Or are the majority of the values outside +/- 1.5?
  • Check calibration records and determine if recalibration is required.
  • Check for a change in mean QC values between lots or shipments of reagents/kits.
  • If participating in a multi-site QC comparison program, check whether your QC means differ from other users of the same lot of QC material.
  • Check all is current and acceptable with QC, calibrations, and preventative maintenance.

Corrective Action

  • Ensure calibration is performed and acceptable after changing lots or shipments of reagents/kits.
  • Verify calibrator values are correctly entered into the instrument.
  • Consider establishing an in-house procedure for validation of manufacturer-provided calibrator values.
  • If calibration drifts or loss of precision is observed over the life of the reagent, consider more frequent calibration, smaller lots/shipments of reagent, or storing the reagent off instrument for a low volume test.
  • If recalibration is necessary for corrective action, rerun any remaining stable PT samples and evaluate against acceptable results.
  • Work with your instrument manufacturer if necessary.

Suspected Cause

    • Reportable Range Issue

Troubleshooting Action

  • Check if previous failure or systemic bias was observed in PT surveys for high or low samples. If the same instrument/reagent had previous problems at high or low values, check to see that the reportable range has been properly determined.

Corrective Action

  • If the reportable range is improperly determined, use reference materials to re-establish the range. If necessary, contact manufacturer for help with troubleshooting instrument/reagent problems.

Suspected Cause

Specimen Integrity

    • Excessive hemolysis (whole blood samples)
    • Excessive precipitate/turbidity/bacterial contamination
    • Matrix Effect (not compatible with method)

Troubleshooting Action

  • Check that samples were in good condition before testing.
  • Check that you have the correct PT module for your analyzer.

Corrective Action

  • Create a process to note and record the condition of the samples upon their arrival.
  • Contact PT provider regarding sample issues.
  • Request replacements if needed. Sample replacement is subject to availability.
  •  

Suspected Cause

Your results show random fluctuations.

Troubleshooting Action

  • The assumption of random error can only be made when all other potential sources of error have been ruled out.
  • May be resolved by retesting the PT sample.

Corrective Action

  • Document as random error or normal statistical variation.